sputter:pellet [2020/07/22 14:41]
jacobs
sputter:pellet [2020/07/27 14:08] (current)
jacobs
Line 4: Line 4:
  
    - Obtain proper target. ​ Wear clean gloves when handling target and target holder. Determine if target is new or used. Used will have a blackened and scared tip.    - Obtain proper target. ​ Wear clean gloves when handling target and target holder. Determine if target is new or used. Used will have a blackened and scared tip.
 +   - Also, determine the proper Buncher Tube setting if you are changing beam species. ​ This will be your only chance to set it.
    - Turn off DEL HIGH VOLTAGE POWER SUPPLY ​ (Main Extractor Power) ​    - Turn off DEL HIGH VOLTAGE POWER SUPPLY ​ (Main Extractor Power) ​
    - Record present value and decrease Cesium Reservoir Current by 50%, usually to about 0.125 amp.    - Record present value and decrease Cesium Reservoir Current by 50%, usually to about 0.125 amp.