playground:powerout

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Last revision Both sides next revision
playground:powerout [2012/02/15 12:52]
jacobs
playground:powerout [2018/01/12 14:58]
jacobs
Line 1: Line 1:
 **Power Failure** **Power Failure**
  
-A power outage lasting longer than 1 minute ​can not be recovered from without the attention of Accelerator staff and will take a minimum of 4 hours before operations can resume.+A power outage lasting longer than 3 minutes will begin to create adverse conditions that can not be recovered from without the attention of Accelerator staff and will take a minimum of 4 hours before operations can resume.  Be Aware: ​ A power failure, even a brief one, will shut down the accelerator,​ sources, and gate valves. ​ Computers and alarms in the control room, and some lights, will continue to function on emergency power from the generator on the roof if it starts properly. ​ You can tell if the power has come back on by the presence of the normal overhead lights.
  
 The Operator should take the following actions after power is restored: The Operator should take the following actions after power is restored:
  
-1)  Silence any alarms that have activated +1)     ​Silence any alarms that have activated, E1/E2, shutdown module, tritium monitor.
-2)  Notify Accelerator Staff of the situation +
-3)  Re-activate the cooling water loop pumps. The controls are located at the entry to the source area. (Ground the high voltage cage). Push each start button. The source cooling water must also be re-activated. This control is on the side of the cabinet that is against the wall by the sputter source. You may need to hold the “on” switch for several seconds before the pump will remain on+
  
-{{playground:​im001091s.jpg|}} {{playground:​im001093s.jpg|}}+2)  Note how long the power was out.
  
-4)  ​Check the status of the vault cryopump compressors. There is one in the high energy pumping station and one next to the low energy pumping station. If they are running, turn them off.  Red toggle switch on the front of the HE compressor, yellow breaker on the rear of the LE compressor.+3)  ​Notify Accelerator Staff of the situation
  
-{{playground:​im001094s.jpg|}} {{playground:​im003474s.jpg|}}+4)  Check the cooling water loop pumpsThe controls are located at the entry to the source area. (Ground the high voltage cage). If the green lights are still on no action is required. ​ If the green lights are off push each start button
  
-5)  Proceed to the mechanical room. Stand at the yellow circle painted on the floor in front of the main breaker panel. Push the reset and start buttons on panels A, B, & C. This will activate 2 air handlers, and the water pump behind you+{{playground:​im001091s.jpg|}} ​
  
-{{playground:​im001097s.jpg|}} {{playground:​im001096s.jpg|}}+5)  The source cooling water must also be re-activatedThis control is on the side of the cabinet that is against the wall by the sputter source. You may need to hold the “on” switch for several seconds before the pump will remain on
  
-6)  Return to the Control Room+{{playground:​im001093s.jpg|}}
  
-7)  ​Set all electrostatic devices ​to zero or neutral, ​and then turn off+6)  ​Check the status of the vault cryopump compressors. There is one next to the LE and HE pumping stations. If they are running, ​turn them off.  Yellow breaker on the rear of the LE compressor. ​ If your beam line has any manual pump gates valves, close them, and shut down your cryopumps if applicable.
  
-8)  On the control computer, go to the Restart tab and select the Settings Logger+ {{playground:​im003474s.jpg|}}
  
-9)  ​Set the TPS and Charging Controller ​to the Initial Condition values+7)  ​No action is required in the upstairs mechanical room.  HVAC systems should return ​to normal within 5 minutes of power restoration.
  
-10Check the status ​of the fire alarm and take action as required+8 ​If ​the accelerator was running when the power went out, set all of its controls to the initial "​off"​ settings.
  
 +9)  Check the status of the magnets. Each magnet control module should be showing a red light for standby or off condition. ​ Magnet current meters need to be showing zero amps.
 +
 +10)  On the control computer, go to the Restart tab and select "​restart all"
  
  
playground/powerout.txt · Last modified: 2020/07/12 13:49 by carter