-  ​- No Operator will intentionally bypass a safety interlock. +**Testing Policy & Interlock Rules** 
-  - Discovery of a failed safety interlock will suspend all operations until that failure can be repaired.+ 
 +  - All safety and warning devices, including interlocks, shall be checked for proper operation at intervals not to exceed three months. ​ Results of such tests shall be maintained in the Testing, Calibration,​ and Shutdown Book. 
 +    - Such testing shall be the responsibility of Accelerator Staff. 
 +    - Interlock Testing: ​ Testing of any one device will be sufficient to demonstrate proper operation of that device'​s interlock circuit. ​ A different device on the interlock circuit will be tested at every three month interval. 
 +    - All audible and visual indicators will be checked for proper operation during each test. 
 +    - **In the event the safety equipment is operating improperly, it shall be immediately reported to the RSO/IRRP and shall be repaired or replaced and never bypassed so as to enable radiation production.** 
 +  - Calibration of fixed and portable monitoring equipment shall be the responsibility of the RSO/IRRP. 
 +    - Calibrations shall be at intervals not to exceed one year and after each servicing or repair. 
 +    - Records of calibrations shall be maintained in the Testing, Calibration,​ and Shutdown Book. 
 +    - **Portable monitoring equipment shall be tested for proper operation daily, prior to use, when in use.** 
 +  - **A safety interlock shall not be used to turn off the Accelerator and/or beam, except in an emergency, or when testing the safety interlock. ​ Routine shutdown shall be done following normal operating procedures and normal controls.** 
 +  ​**No Operator will intentionally bypass a safety interlock.** 
 +  - **Discovery of a failed safety interlock ​or warning device ​will suspend all operations until that failure can be repaired.**
   - If, for any reason, the TALC finds it necessary to intentionally bypass a safety interlock this action shall be:   - If, for any reason, the TALC finds it necessary to intentionally bypass a safety interlock this action shall be:
-  * Authorized by the OU Radiation Safety Officer; +    - Authorized by the OU Radiation Safety Officer 
-  ​* ​Recorded in the Electronic Logbook and emailed to all recipients; +    ​- ​Recorded in the Electronic Logbook and emailed to all recipients 
-  ​* ​Posted as a written notice at the Accelerator Control Consoleand +    ​- ​Posted as a written notice at the Accelerator Control Consoleand 
-  ​* ​Terminated as soon as possible.+    ​- ​Terminated as soon as possible. 
  