This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | Next revision Both sides next revision | ||
playground:integrator [2012/03/20 15:06] carter |
playground:integrator [2012/03/20 15:17] carter |
||
---|---|---|---|
Line 4: | Line 4: | ||
* a. The digital display reads out the total integrated charge in Coulombs * Multiplier (including the decimal point). The pre-scaler value (h) is automatically included in the display by moving the decimal point in the display readout. For example, if the digital display reads 1234.56 and the range is 10**-6 the charge is 1234.56 uC. | * a. The digital display reads out the total integrated charge in Coulombs * Multiplier (including the decimal point). The pre-scaler value (h) is automatically included in the display by moving the decimal point in the display readout. For example, if the digital display reads 1234.56 and the range is 10**-6 the charge is 1234.56 uC. | ||
+ | |||
* b. This switch sets the 10**-X range value. | * b. This switch sets the 10**-X range value. | ||
+ | |||
* c. This is the instantaneous beam current value. Full scale is the 1.0 * 10**(range switch) Amps full scale. | * c. This is the instantaneous beam current value. Full scale is the 1.0 * 10**(range switch) Amps full scale. | ||
+ | |||
* d. This switch is used to expand the full scale indication of the beam current on the meter only. It has no effect on the total integrated charge value. | * d. This switch is used to expand the full scale indication of the beam current on the meter only. It has no effect on the total integrated charge value. | ||
* e. This switch selects the polarity of the input current. | * e. This switch selects the polarity of the input current. |