Changing Terminal Energy

1. Switch OVERVOLTAGE PROTECTION to DISABLE.

2. Switch the control mode to GVM.

3. Turn the CONTROL GAIN to zero.

4. To go to a higher energy move the CORONA PROBE POSITION “OUT” while raising the CHARGING VOLTAGE. NEVER allow the GRID VOLTAGE to go more negative than -10 VDC during this operation.

5. To go to a lower energy move the CORONA PROBE POSITION “IN” while lowering the CHARGING VOLTAGE. NEVER allow the GRID VOLTAGE to go more negative than -10 VDC during this operation.

6. Switch POWER SUPPLIES to OFF.

7. Turn the TERMINAL VOLTAGE dial to zero.