- Switch OVERVOLTAGE PROTECTION to DISABLE. | - Switch OVERVOLTAGE PROTECTION to DISABLE. | ||
- Switch the control mode to GVM. | - Switch the control mode to GVM. | ||
- | - Turn the CONTROL GAIN to zero. | + | - Slowly turn the CONTROL GAIN to zero while watching the grid voltage. If the grid voltage rises above 10 volts, reduce the CHARGING VOLTAGE to keep the grid voltage less than 10 volts, then continue reducing the CONTROL GAIN to zero. |
- To go to a higher energy move the CORONA PROBE POSITION "OUT" while raising the CHARGING VOLTAGE. **__NEVER__** allow the GRID VOLTAGE to go more negative than -10 VDC during this operation. | - To go to a higher energy move the CORONA PROBE POSITION "OUT" while raising the CHARGING VOLTAGE. **__NEVER__** allow the GRID VOLTAGE to go more negative than -10 VDC during this operation. | ||
- To go to a lower energy move the CORONA PROBE POSITION "IN" while lowering the CHARGING VOLTAGE. **__NEVER__** allow the GRID VOLTAGE to go more negative than -10 VDC during this operation. | - To go to a lower energy move the CORONA PROBE POSITION "IN" while lowering the CHARGING VOLTAGE. **__NEVER__** allow the GRID VOLTAGE to go more negative than -10 VDC during this operation. |