This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | Last revision Both sides next revision | ||
playground:alphatross [2022/01/31 11:57] leblanc Typos |
playground:alphatross [2022/06/03 11:36] leblanc Remove waiting step, add camera position, fix capitalization. |
||
---|---|---|---|
Line 16: | Line 16: | ||
== In the Control Room == | == In the Control Room == | ||
- | - Source camera must have a view of the faraday screen on the blue box of the source head. | + | - Source camera must have a view of the Faraday screen on the blue box of the source head. |
- To gain control of the source camera: On the joystick unit press "2" then "CAM" to specify the camera. | - To gain control of the source camera: On the joystick unit press "2" then "CAM" to specify the camera. | ||
- | - Use the joystick and "TELE" or "WIDE" to create a view of interest. | + | - Use the joystick and "TELE" or "WIDE" to create a view of interest. Pressing "6" followed by "Call" will move and zoom the camera appropriately. |
- If not already present, launch the Alphatross software (Version 6.0) on the source computer. | - If not already present, launch the Alphatross software (Version 6.0) on the source computer. | ||
- Press the "on" button for Water Cooling. The water cooling system: | - Press the "on" button for Water Cooling. The water cooling system: | ||
Line 45: | Line 45: | ||
- Source is still grounded. | - Source is still grounded. | ||
- Source is dirty and requires service. | - Source is dirty and requires service. | ||
- | - Set Extractor Voltage to 5kV. | + | - Set Extractor Voltage to 5kV. |
- | - Set Bias Voltage to a value that gives desired injection energy. Typical values are at least 53keV. | + | - Set Bias Voltage to a value that gives desired injection energy. Typical values are at least 53keV. |
- | - Injection energy is displayed in the upper window. | + | - Injection energy is displayed in the upper window. |
- | - Injection energy is the sum of Probe, Extractor, and Bias voltages. | + | - Injection energy is the sum of Probe, Extractor, and Bias voltages. |
- | - Beam yields improve above 30 keV. | + | - Beam yields improve above 30 keV. |
- | - Set the Focus Voltage to 5.5 kV | + | - Set the Focus Voltage to 5.5 kV |
- | - Set the Faraday cup meter to position 7 (alphatross cup). | + | - Set the Faraday cup meter to position 7 (Alphatross cup). |
- | - Tune for maximum __stable__ beam. | + | - Tune for maximum __stable__ beam. |
- | - Primary adjustment at this stage seems to be focus voltage. | + | - Primary adjustment at this stage seems to be focus voltage. |
- | - Current may be on the order of 30 μA. | + | - Current may be on the order of 30 μA. |
- | + | ||
- | ==**DO NOT PROCEED UNTIL CHAMBER TEMPERATURE EXCEEDS 39°C**== | + | |
- Set the Oven Voltage to 80 V. | - Set the Oven Voltage to 80 V. | ||
- Set Einzel Lens to 20 kV. | - Set Einzel Lens to 20 kV. |