Line 3: Line 3:
 ** Duoplasmatron Instructions ** ** Duoplasmatron Instructions **
  
-Please study the diagram.  +**Please study the diagram ​below.** 
-{{playground:​duoplasmatron.jpg|}}+ 
 +{{playground:​duo.jpg|}}
  
  
Line 56: Line 57:
 5. Turn on the focus supply, it should be set for ~2 kV. 5. Turn on the focus supply, it should be set for ~2 kV.
  
-6. Turn up the sodium heater current to 0.55 A, and the filament current up slowly, over 30 seconds, until the arc starts.+6. Turn up the sodium heater current to 0.55 A.
  
-7. When the arc starts, ​70 A for an old filament, ​90 A for a new filament, observe the Arc Voltage and keep it at 150 V.+7. Slowly turn up the Filament Current until the arc starts.  ​When the arc starts, ​60 A for an old filament, ​70 a for a new filament, observe the Arc Voltage and keep it at 150 V.
  
 8. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. 8. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V.
Line 134: Line 135:
 10. Increase the sodium heater current by ~0.05 A and the filament current by ~10 A every twenty minutes. Make sure the pressure does not rise over 1x10<​sup>​-5</​sup>​ torr, if it does, do not proceed further until it has fallen below this value. 10. Increase the sodium heater current by ~0.05 A and the filament current by ~10 A every twenty minutes. Make sure the pressure does not rise over 1x10<​sup>​-5</​sup>​ torr, if it does, do not proceed further until it has fallen below this value.
  
-11. When the arc starts, ​70 A for a dirty source or old filament, ​90 A for a new filament ​in a clean source, observe the arc voltage ​and keep it at 150 V.+11. Slowly turn up the Filament Current until the arc starts.  ​When the arc starts, ​60 A for an old filament, ​70 a for a new filament, observe the Arc Voltage ​and keep it at 150 V. 
  
 12. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the arc voltage at 150 V. 12. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the arc voltage at 150 V.