** Duoplasmatron Instructions ** | ** Duoplasmatron Instructions ** | ||
- | Please study the diagram. | + | **Please study the diagram below.** |
- | {{playground:duoplasmatron.jpg|}} | + | |
+ | {{playground:duoplasmatron_big.jpg|}} | ||
Line 56: | Line 57: | ||
5. Turn on the focus supply, it should be set for ~2 kV. | 5. Turn on the focus supply, it should be set for ~2 kV. | ||
- | 6. Turn up the sodium heater current to 0.55 A, and the filament current up slowly, over 30 seconds, until the arc starts. | + | 6. Turn up the sodium heater current to 0.55 A. |
- | 7. When the arc starts, 70 A for an old filament, 90 A for a new filament, observe the Arc Voltage and keep it at 150 V. | + | 7. Slowly turn up the Filament Current until the arc starts. When the arc starts, 60 A for an old filament, 70 a for a new filament, observe the Arc Voltage and keep it at 150 V. |
8. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. | 8. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. | ||
- | 9. Increase the sodium heater current until it reaches 0.6 A, **or whatever was used the last time** the source was run – check the log. | + | 9. Increase the sodium heater current until it reaches up to 0.6 A, **or whatever was used the last time** the source was run – check the log. |
10. Set the inflection magnet to about 0.8 amps and LEFT deflection. | 10. Set the inflection magnet to about 0.8 amps and LEFT deflection. | ||
Line 77: | Line 78: | ||
* Focus: 2 kV @ 1.5 mA | * Focus: 2 kV @ 1.5 mA | ||
* Magnet : 0.2 - 1.4A | * Magnet : 0.2 - 1.4A | ||
- | * Heater: 0.65 A | + | * Heater: 0.58 A |
* Arc Voltage: 150 V | * Arc Voltage: 150 V | ||
* Filament Current: 70 - 90 amps | * Filament Current: 70 - 90 amps | ||
Line 86: | Line 87: | ||
* 2Y: 5.00 | * 2Y: 5.00 | ||
* 2X: 5.00 | * 2X: 5.00 | ||
- | + | * Box Vacuum: 2.0x10E<sup>-6</sup> torr | |
- | Box Vacuum: 2.0x10E<sup>-6</sup> torr | + | |
** Starting from cold ** | ** Starting from cold ** | ||
Line 135: | Line 135: | ||
10. Increase the sodium heater current by ~0.05 A and the filament current by ~10 A every twenty minutes. Make sure the pressure does not rise over 1x10<sup>-5</sup> torr, if it does, do not proceed further until it has fallen below this value. | 10. Increase the sodium heater current by ~0.05 A and the filament current by ~10 A every twenty minutes. Make sure the pressure does not rise over 1x10<sup>-5</sup> torr, if it does, do not proceed further until it has fallen below this value. | ||
- | 11. When the arc starts, 70 A for a dirty source or old filament, 90 A for a new filament in a clean source, observe the arc voltage and keep it at 150 V. | + | 11. Slowly turn up the Filament Current until the arc starts. When the arc starts, 60 A for an old filament, 70 a for a new filament, observe the Arc Voltage and keep it at 150 V. |
12. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the arc voltage at 150 V. | 12. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the arc voltage at 150 V. | ||
- | 13. Increase the heater current in 0.05 A steps until it reaches 0.6 A. | + | 13. Increase the heater current in 0.05 A steps until it reaches 0.57 A. |
14. Set inflection magnet to ~0.8 amps and LEFT deflection. | 14. Set inflection magnet to ~0.8 amps and LEFT deflection. | ||
Line 153: | Line 154: | ||
17. Typical Settings: | 17. Typical Settings: | ||
* Extraction : 12 kV @ < 1 mA | * Extraction : 12 kV @ < 1 mA | ||
- | * Attachment: 13 kV @ 1 mA | + | * Exchange Canal: 13 kV @ 1 mA |
* Focus: 2 kV @ 1.5 mA | * Focus: 2 kV @ 1.5 mA | ||
- | * Magnet : 0.2 A | + | * Magnet : 0.2 - 1.4A |
- | * Heater: 0.67 A | + | * Heater: 0.58 A |
- | * Arc voltage: 150 V | + | * Arc Voltage: 150 V |
* Filament Current: 70 - 90 amps | * Filament Current: 70 - 90 amps | ||
* Arc Current: 2.5 amps | * Arc Current: 2.5 amps | ||
Line 173: | Line 174: | ||
1. Typical Settings: | 1. Typical Settings: | ||
* Extraction : 12 kV @ < 1 mA | * Extraction : 12 kV @ < 1 mA | ||
- | * Attachment: 13 kV @ 1 mA | + | * Exchange Canal: 13 kV @ 1 mA |
* Focus: 2 kV @ 1.5 mA | * Focus: 2 kV @ 1.5 mA | ||
- | * Magnet : 0.2 A | + | * Magnet : 0.2 - 1.4A |
- | * Sodium Heater: 0.55 A | + | * Heater: 0.55 A |
* Arc Voltage: 150 V | * Arc Voltage: 150 V | ||
* Filament Current: 70 - 90 amps | * Filament Current: 70 - 90 amps | ||
- | * Arc Current: 0.75 amps | + | * Arc Current: 2.5 amps |
* Source lens: 2 kV | * Source lens: 2 kV | ||
* Lens A: <5 kV | * Lens A: <5 kV | ||
Line 185: | Line 186: | ||
* 2Y: 5.00 | * 2Y: 5.00 | ||
* 2X: 5.00 | * 2X: 5.00 | ||
- | *Box Vacuum: 2.0x10E<sup>-6</sup> torr | + | * Box Vacuum: 2.0x10E<sup>-6</sup> torr |
Line 205: | Line 206: | ||
1. Increase the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. | 1. Increase the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. | ||
- | 2. Increase the sodium heater current until it reaches 0.65 A, or whatever was used the last time the source was run – check the log. | + | 2. Increase the sodium heater current until it reaches up to 0.6 A, **or whatever was used the last time** the source was run – check the log. |
3. Set the inflection magnet to 0.8 amps and LEFT deflection. | 3. Set the inflection magnet to 0.8 amps and LEFT deflection. | ||
Line 217: | Line 218: | ||
5. Typical Settings: | 5. Typical Settings: | ||
* Extraction : 12 kV @ < 1 mA | * Extraction : 12 kV @ < 1 mA | ||
- | * Attachment: 13 kV @ 1 mA | + | * Exchange Canal: 13 kV @ 1 mA |
* Focus: 2 kV @ 1.5 mA | * Focus: 2 kV @ 1.5 mA | ||
- | * Magnet : 0.7 A | + | * Magnet : 0.2 - 1.4A |
- | * Heater: 0.67 A | + | * Heater: 0.58 A |
- | * Arc voltage: 150 V | + | * Arc Voltage: 150 V |
* Filament Current: 70 - 90 amps | * Filament Current: 70 - 90 amps | ||
* Arc Current: 2.5 amps | * Arc Current: 2.5 amps | ||
Line 229: | Line 230: | ||
* 2Y: 5.00 | * 2Y: 5.00 | ||
* 2X: 5.00 | * 2X: 5.00 | ||
- | *Box Vacuum: 2.0x10E<sup>-6</sup> torr | + | * Box Vacuum: 2.0x10E<sup>-6</sup> torr |
** Duoplasmatron Warm Shutdown Procedure ** | ** Duoplasmatron Warm Shutdown Procedure ** | ||
Line 243: | Line 244: | ||
4. Turn off the deflection supplies and run lenses down to zero. | 4. Turn off the deflection supplies and run lenses down to zero. | ||
- | 5. Turn down the source gas until the pressure on the source tc gauge is ~10 microns. Stop pressing the button when the pressure is 100 microns to allow the tc gauge to catch up. The pressure response is delayed. Do not over-drive the source gas control as it could damage the leak valve. | + | 5. Turn down the source gas until the pressure on the source tc gauge is ~10 microns. The pressure response is delayed, be patient. Do not over-drive the source gas control as it could damage the leak valve. |
** Duoplasmatron Shut Down Completely ** | ** Duoplasmatron Shut Down Completely ** | ||
Line 260: | Line 261: | ||
5. Turn off the deflection supplies and run lenses down to zero. | 5. Turn off the deflection supplies and run lenses down to zero. | ||
- | 6. Turn down the source gas until the pressure on the source tc gauge is ~10 microns.� Stop pressing the button when the pressure is 100 microns to allow the tc gauge to catch up. The pressure response is delayed.� Do not over-drive the source gas control as it could damage the leak valve. | + | 6. Turn down the source gas until the pressure on the source tc gauge is ~10 microns. The pressure response is delayed, be patient. Do not over-drive the source gas control as it could damage the leak valve. |
7. After 10 minutes to allow cooling water to circulate, turn control power key off on source control panel. | 7. After 10 minutes to allow cooling water to circulate, turn control power key off on source control panel. | ||
Line 282: | Line 283: | ||
This assumes the source is running on <sup>4</sup>He already and that the regulator on the <sup>4</sup>He bottle is set for no more than 5 psig. | This assumes the source is running on <sup>4</sup>He already and that the regulator on the <sup>4</sup>He bottle is set for no more than 5 psig. | ||
- | Turn down the source gas until the pressure on the source tc gauge is ~10 microns.� Stop pressing the button when the pressure is 100 microns to allow the tc gauge to catch up.� The pressure response is delayed.� Do not over-drive the source gas control as it could damage the leak valve. | + | Turn the Filament Current and Arc Voltage to zero |
+ | |||
+ | Turn down the source gas until the pressure on the source tc gauge is ~10 microns. The pressure response is delayed, be patient. Do not over-drive the source gas control as it could damage the leak valve. | ||
Go out to the vault and make sure a <sup>3</sup>He bottle is ready for use. It MUST have its main valve closed at this point. | Go out to the vault and make sure a <sup>3</sup>He bottle is ready for use. It MUST have its main valve closed at this point. | ||
Line 296: | Line 299: | ||
Close the manifold valve to the rough pump and turn off the pump. | Close the manifold valve to the rough pump and turn off the pump. | ||
- | Open the 3He bottle main valve, regulator should go to ~ 5 psig.� Record the bottle pressure on the tag. | + | Open the 3He bottle main valve, regulator should go to ~ 5 psig. Record the bottle pressure on the tag. |
- | At the control console, turn up the source gas until the pressure on the source tc gauge is ~150 microns.� Flip the tag on the panel around to show 3He. | + | At the control console, turn up the source gas until the pressure on the source tc gauge is ~300 microns. Flip the tag on the panel around to show 3He. |
Reverse this procedure to return to 4He operation, except don’t rough beyond the 4He manifold valve (don’t rough line and regulator) when you have finished using 3He. | Reverse this procedure to return to 4He operation, except don’t rough beyond the 4He manifold valve (don’t rough line and regulator) when you have finished using 3He. | ||
Line 329: | Line 332: | ||
Telephone: (740)593-0933 | Telephone: (740)593-0933 | ||
E-Mail: MASSEY@OHIOU.EDU | E-Mail: MASSEY@OHIOU.EDU | ||
- | Last modified on 06/20/06 | + | Last modified on 3/16/2012 TNM |