Line 133: | Line 133: | ||
A "HIGH RAD MODE" switch is located on the Faraday Cup Control Module. When this switch is in the "OFF" position the LE cup will lift instantaneously and the audible and visual warning will be active for 15 seconds. When this switch is in the "ON" position there will be a 15 second delay between pushing the LE Cup "OUT" button and the actual lifting of the LE Cup, as well as the 15 second audible/visual warning. | A "HIGH RAD MODE" switch is located on the Faraday Cup Control Module. When this switch is in the "OFF" position the LE cup will lift instantaneously and the audible and visual warning will be active for 15 seconds. When this switch is in the "ON" position there will be a 15 second delay between pushing the LE Cup "OUT" button and the actual lifting of the LE Cup, as well as the 15 second audible/visual warning. | ||
- | The Operator shall place the HIGH RAD MODE switch in the "ON" position when: | + | **The Operator shall place the HIGH RAD MODE switch in the "ON" position when:** |
* The experiment conditions are known to produce measurable radiation | * The experiment conditions are known to produce measurable radiation | ||
Line 139: | Line 139: | ||
* Any time there is uncertainty in the radiation that will result from the experiment conditions | * Any time there is uncertainty in the radiation that will result from the experiment conditions | ||
- | The Operator may place the HIGH RAD MODE switch in the "OFF" position when: | + | **The Operator may place the HIGH RAD MODE switch in the "OFF" position when:** |
* The experiment conditions are known to NOT produce measurable radiation | * The experiment conditions are known to NOT produce measurable radiation | ||
Line 170: | Line 170: | ||
The Tritium monitor is normally left on the x1 scale with the alarm level set at the maximum of the range (~10 microcuries/meters3). The maximum allowable level is 20 microcuries/meters3. | The Tritium monitor is normally left on the x1 scale with the alarm level set at the maximum of the range (~10 microcuries/meters3). The maximum allowable level is 20 microcuries/meters3. | ||
- | **If the Tritium monitor is alarming:** | + | **Actions if the Tritium monitor is alarming:** |
////-Check that there is air flow through the ion chamber. The Tritium Monitor will alarm if the pump is not on//// | ////-Check that there is air flow through the ion chamber. The Tritium Monitor will alarm if the pump is not on//// |