playground:alarms [2016/05/04 09:53]
jacobs [Pelletron AC Power Center: Located in the Vault by the sputter source gate.]
playground:alarms [2016/05/04 10:36] (current)
jacobs [Tritium Monitor:]
Line 109: Line 109:
  
 {{playground:​im003639.jpg|}} {{playground:​im003639.jpg|}}
 +{{playground:​emopanel3.jpg|}}
  
-  ​- Identify which EMO was pushed and investigate the situation that led to the Emergency Stop event.+**Actions:​** 
 +  ​- Identify which EMO was pushed ​(see the Emergency Off Annunciator Panel) ​and investigate the situation that led to the Emergency Stop event.
   - Follow Section VIII of the Radiation Safety Handbook if there was any possibility of exposure.   - Follow Section VIII of the Radiation Safety Handbook if there was any possibility of exposure.
   - Otherwise, provide a detailed written account of the event in the Electronic Logbook. ​ Describe who was involved, which button was pushed, why it was pushed, and describe the course of events thast led to the Emergency Stop.   - Otherwise, provide a detailed written account of the event in the Electronic Logbook. ​ Describe who was involved, which button was pushed, why it was pushed, and describe the course of events thast led to the Emergency Stop.
-  - Reconfigue ​the TPS and CHARGING CONTROLS as directed in the Accelerator Shutdown procedure.+  - Reconfigure ​the TPS and CHARGING CONTROLS as directed in the Accelerator Shutdown procedure.
   - Reset the activated EMO.  Turn the red knob in the direction indicated by the arrows. ​ The knob will pop outwards when it releases.   - Reset the activated EMO.  Turn the red knob in the direction indicated by the arrows. ​ The knob will pop outwards when it releases.
   - Go to the Vault and reset the Pelletron AC Power Center. ​ You may hear faraday cups operate when you do this.   - Go to the Vault and reset the Pelletron AC Power Center. ​ You may hear faraday cups operate when you do this.
   - Go to the Control Room and place all faraday cups in the "​IN"​ position.   - Go to the Control Room and place all faraday cups in the "​IN"​ position.
   - Follow the Accelerator Startup Procedure and resume operations.   - Follow the Accelerator Startup Procedure and resume operations.
- 
  
    
Line 132: Line 133:
 A "HIGH RAD MODE" switch is located on the Faraday Cup Control Module. ​ When this switch is in the "​OFF"​ position the LE cup will lift instantaneously and the audible and visual warning will be active for 15 seconds. ​ When this switch is in the "​ON"​ position there will be a 15 second delay between pushing the LE Cup "​OUT"​ button and the actual lifting of the LE Cup, as well as the 15 second audible/​visual warning. A "HIGH RAD MODE" switch is located on the Faraday Cup Control Module. ​ When this switch is in the "​OFF"​ position the LE cup will lift instantaneously and the audible and visual warning will be active for 15 seconds. ​ When this switch is in the "​ON"​ position there will be a 15 second delay between pushing the LE Cup "​OUT"​ button and the actual lifting of the LE Cup, as well as the 15 second audible/​visual warning.
  
- The Operator shall place the HIGH RAD MODE switch in the "​ON"​ position when:+ **The Operator shall place the HIGH RAD MODE switch in the "​ON"​ position when:**
  
   * The experiment conditions are known to produce measurable radiation   * The experiment conditions are known to produce measurable radiation
Line 138: Line 139:
   * Any time there is uncertainty in the radiation that will result from the experiment conditions   * Any time there is uncertainty in the radiation that will result from the experiment conditions
  
- The Operator may place the HIGH RAD MODE switch in the "​OFF"​ position when:+ **The Operator may place the HIGH RAD MODE switch in the "​OFF"​ position when:**
  
   * The experiment conditions are known to NOT produce measurable radiation   * The experiment conditions are known to NOT produce measurable radiation
Line 169: Line 170:
 The Tritium monitor is normally left on the x1 scale with the alarm level set at the maximum of the range (~10  microcuries/​meters3). ​   The maximum allowable level is 20 microcuries/​meters3. The Tritium monitor is normally left on the x1 scale with the alarm level set at the maximum of the range (~10  microcuries/​meters3). ​   The maximum allowable level is 20 microcuries/​meters3.
  
-**If the Tritium monitor is alarming:**+**Actions if the Tritium monitor is alarming:**
  
 ////-Check that there is air flow through the ion chamber. ​ The Tritium Monitor will alarm if the pump is not on//// ////-Check that there is air flow through the ion chamber. ​ The Tritium Monitor will alarm if the pump is not on////