Line 25: Line 25:
 ==== Monitoring Equipment ==== ==== Monitoring Equipment ====
  
->**Below this point is the nitty-gritty details on the oxygen monitors, wiring, manuals, etc.  Operators shouldn'​t need to refer to this information,​ but it is made available for accelerator staff.**+>**Below this point is the nitty-gritty details on the oxygen monitors, wiring, manuals, etc.  Operators shouldn'​t need to refer to this information,​ but it is made available for all.**
  
 === Equipment Info === === Equipment Info ===
Line 35: Line 35:
 If low oxygen is detected, an alarm will sound, and the area should be evacuated, and accelerator staff contacted. If low oxygen is detected, an alarm will sound, and the area should be evacuated, and accelerator staff contacted.
  
-Correspondence with AMI can be found at: {{::​ami_sensor_replacement.pdf|}} and also at {{::​ami_2021-06-22.pdf|}}+Correspondence with AMI can be found at: {{::​ami_sensor_replacement.pdf|}} and also at {{::​ami_2021-06-22.pdf|}}.
  
   * Sensor data sheet: ​   * Sensor data sheet: ​
Line 48: Line 48:
  
 == Sensor Info == == Sensor Info ==
-The monitors are equipped with remote probes, to allow placement closer to the floor, since the risk is primarily from heavier than air gasses. ​ The Ziurconium Oxide sensors used in these devices are expected to have a 10 year service life.  The sensor is located in the remote probe, and could theoretically be sent back for repair separately from the head unit.  AMI recommends returning both the head unit and sensor together. ​ Sensors are approaching end of life when their span factor ​(checked through Command Center Software) ​is approaching 1000.+The monitors are equipped with remote probes, to allow placement closer to the floor, since the risk is primarily from heavier than air gasses. ​ The Ziurconium Oxide sensors used in these devices are expected to have a 10 year service life.  The sensor is located in the remote probe, and could theoretically be sent back for repair separately from the head unit.  AMI recommends returning both the head unit and sensor together. ​ Sensors are approaching end of life when their span factor is approaching 1000.  The span factor can be checked through the command center software, or by pressing the "​up"​ button on the front panel.
  
 = Calibration Info = = Calibration Info =
oxygen_monitors.1624466611.txt.gz ยท Last modified: 2021/06/23 12:43 by leblanc