Line 25: Line 25:
 ** Startup and Normal Operation (in control room) ** ** Startup and Normal Operation (in control room) **
  
-  - Turn on and set the Del main extractor supply to the desired voltage for the particular buncher setting. +  - Turn on and set the **Del main extractor** supply to the desired voltage for the particular buncher setting. 
-  - Slowly turn up the ionizer heater. ​  The ionizer operates at 21 amperes. It should take about 2 minutes to reach 21 A from the standby value of 10 A. Keep in mind that the amperage ​will fall slightly as the ionizer gets warmer. ​ Keep adjusting the current until the ionizer is stable at 21 A. +  - Slowly turn up the ionizer heater. ​  The ionizer operates at 21 amperes. It should take about 2 minutes to reach 21 A from the standby value of 10 A. Keep in mind that the current ​will fall slightly as the ionizer gets warmer. ​ Keep adjusting the current until the ionizer is stable at 21 A. 
-  - Turn up the Cs feed tube heater to 0.450 amperes. ​  +  - Turn up the Cs **feed tube** heater to 0.450 amperes. ​  
-  - Set the cesium reservoir heater to 0.380. This should be enough to give ones of microamps beam for p and d.    A setting of about 0.400 - 0.420 should produce tens of microamps. Cs vaporization starts at a temperature of ~80° Celsius. ​  The maximum reservoir temperature should not exceed ​100° Celsius. The reservoir current should not exceed 0.500 +  - Set the cesium ​**reservoir** heater to 0.285. This should be enough to give ones of microamps beam for p and d.    A setting of about 0.290 - 0.300 should produce tens of microamps. Cs vaporization starts at a temperature of ~80° Celsius. ​  The maximum reservoir temperature should not exceed ​130° Celsius. The reservoir current should not exceed 0.300. 
-  - Set the Glassman extractor power supply to 12.5 KV. Check to be sure the current is less than about 5 milliamps and the "​control ​milliampres" light is NOT lit. High current indicates that the source needs serviced. +  - Set the Glassman ​**extractor** power supply to 12.5 KV. Check to be sure the current is less than about 5 milliamps and the "​control ​milliamperes" light is NOT lit. High current indicates that the source needs to be serviced. 
-  - Set the Glassman target power supply to 6.5 kV. Check to be sure the current is less than about 5 milliamps and the "​control ​milliampres" light is NOT lit. High current indicates that the source needs serviced. +  - Set the Glassman ​**target** power supply to 6.5 kV. Check to be sure the current is less than about 5 milliamps and the "​control ​milliamperes" light is NOT lit. High current indicates that the source needs to be serviced
-  - Tune the beam with the inflection magnet, source lens (usually about 1.5 kV), lens A, and 2Y steerer.+  - Set the Glassman **lens** power supply to 1.5 kV
 +  - Tune the beam with the inflection magnet, source lens (usually about 1.5 kV), einzel ​lens A, and 2Y steerer.  The 2X steerer should be near neutral (5.00 on dial) as X steering is be done by the inflection magnet at this stage.
   - Monitor the target and extractor currents: target current should be around 2.5 milliamps, and no more than 5 mA.   - Monitor the target and extractor currents: target current should be around 2.5 milliamps, and no more than 5 mA.
  
Line 38: Line 39:
   - The source gate valve should be closed for this operation.   - The source gate valve should be closed for this operation.
   - Wait for normal vacuum. ​ Normal is when the needle is in the blue mark on the gauge above the TV in the control room.   - Wait for normal vacuum. ​ Normal is when the needle is in the blue mark on the gauge above the TV in the control room.
-  - Set ionizer to 10A.� This will lead to a rise in the vacuum pressure.+  - Set ionizer to 10A. This will lead to a rise in the vacuum pressure.
   - Wait for vacuum to fall back to the blue mark on the gauge   - Wait for vacuum to fall back to the blue mark on the gauge
-  - Raise the ionizer by an additional� 2-3 A and wait for the vacuum to fall back into the blue range. +  - Raise the ionizer by an additional 2-3 A and wait for the vacuum to fall back into the blue range. 
-  - Repeat this process until the ionizer is at 21 A and the vacuum is in the blue range.+  - Repeat this process until the ionizer is at 21 A and the vacuum is in the blue range, may take 8 hours.
   - Open the source gate valve and carry out the Startup and Normal Operation.   - Open the source gate valve and carry out the Startup and Normal Operation.
  
Line 67: Line 68:
 Telephone: (740)593-0933\\ Telephone: (740)593-0933\\
 E-Mail: massey@ohio.edu\\ E-Mail: massey@ohio.edu\\
-Last modified on 12/06/2006\\+Last modified on 7/30/2010 DKJ\\
  
  
playground/sputter.txt · Last modified: 2023/08/22 07:50 by carter