Line 14: Line 14:
 {{playground:​im001094s.jpg|}}{{playground:​im001095s.jpg|}}{{playground:​im003474s.jpg|}} {{playground:​im001094s.jpg|}}{{playground:​im001095s.jpg|}}{{playground:​im003474s.jpg|}}
  
-3)     ​Return to the sputter source vacuum cabinet. ​ Activate the systems by pushing the buttons indicated by the green arrows, from left to right, Fore Pump, Freon Compressor, Diffusion Pump, and, if those buttons have all turned green, the Baffle (gate) Valve. ​ If the diffusion pump will not come on, check that the source cooling water is turned on. +3) Proceed to the mechanical room.  Stand at the yellow circle painted on the floor in front of the main breaker panel. ​ Push the reset and start buttons on panels A, B, & C.  This will activate 2 air handlers, and the water pump behind you.
- +
-{{playground:​im003473s.jpg|}} +
- +
- +
-4) Proceed to the mechanical room.  Stand at the yellow circle painted on the floor in front of the main breaker panel. ​ Push the reset and start buttons on panels A, B, & C.  This will activate 2 air handlers, and the water pump behind you.+
  
 {{playground:​im001097s.jpg|}}{{playground:​im001096s.jpg|}} {{playground:​im001097s.jpg|}}{{playground:​im001096s.jpg|}}
  
-5) Return to the control room.  Check the “main loop temp” on the alarm panel. ​ If it is greater than 50 the chiller has failed and you will need staff to restart it.  Do not re-start the accelerator,​ and turn down the magnets.+5) Return to the control room.  Check the “main loop temp” on the alarm panel. ​ If it is greater than 58 the chiller has failed and you will need staff to restart it.  Do not re-start the accelerator,​ and turn down the magnets.
  
-6) Return to the vault and check that the 4 cryopumps ( LE line, LE pumping station, HE pumping station, analyzer line) are still cold.  Remember, the green part of the dial is NOT good.  If they have failed, shut down the compressor to begin regeneration.+6) Return to the vault and check that the 4 cryopumps ( LE line, LE pumping station, HE pumping station, analyzer line) are still cold.  Remember, the green part of the dial is NOT good.  Condensation or frost on the cryopump is a sign of failure.  If they have failed, shut down the compressor to begin regeneration. ​ 
  
-7) This process has preserved the vacuum ​and cooling systems. ​ The ion gauges, gates valves and ion sources will need to be carefully re-started to resume operations. ​ Call the lab staff personnel for assistance in these matters.+7) This process has preserved the vacuumcooling, climate control ​systems. ​ The ion gauges, gates valves and ion sources will need to be carefully re-started to resume operations. ​ Call the lab staff personnel for assistance in these matters.
  