 ==== NEC Alphatross RF Source Instructions ==== ==== NEC Alphatross RF Source Instructions ====
    
-** Duoplasmatron Instructions **+THESE ARE DRAFT INSTRUCTIONS WRITTEN REMOTELY. ​ THERE MAY BE OMISSIONS AND INCORRECT NOMENCLATURE,​ USE YOUR OWN JUDGEMENT...
  
-**Please study the diagram below.**+**  ​__START UP__ **
  
-{{playground:​duo.jpg|}}+** In the Vault **
  
 +  - LE Extension vacuum must be good as evidenced by Ion Gauge controller with orange displays on magnet.
 +  - Source vacuum must be good as evidenced by Ion Gauge controller with blue displays in 19" rack.
 +  - Unground alpha source and hang shepherd'​s hook on concrete wall conduits, close gate.
 +  - Go around and unground sputter source, hang shepherd'​s hook on gate, and close gate.
  
 +** In the Control Room **
  
-These instructions were developed by David Ingram, please e-mail any comments or questions about them to him, at ingram@ohio.edu+  ​Source camera must have a view of the faraday screen on the blue box of the source head  
- +    - To gain control of the source cameraOn the joystick unit press "​2"​ then "​CAM" ​to specify the camera
-This page includesDuoplasmatron cold start, warm start, start from standby; shutdown completely, shutdown for warm start, shutdown ​to standby; Trouble shooting; 3He operation+    Use the joystick and "​TELE"​ or "​WIDE" ​to create a view of interest. 
- +  - If not already present, launch the Alphatross software ​on the source ​computer. 
- +  - In prompt succession press the "​on"​ buttons for: Water CoolingSource Powerand He4 Gas
-** Duoplasmatron Start-up ** +    - These actions start a 60 second timer on the RF supply and set initial conditions on the Magnet and Mass Flow Controller. 
- +    ​After 60 seconds ​the RF power starts, gives a reading on the meter above the faraday screen, and hopefully ​the plasma will ignite giving an intense white glow.  It is bright enough to illuminate the faraday screen ​if not looking directly ​at it in the mirror
-You need to determine what the state of the source is before you do anythingDepending ​on the electrical and vacuum status ​the source ​will be starting from coldwarmor standby+  - Plasma or not, raise the Magnet ​to A, maybe the plasma will ignite
- +    - Monitor ​the Ion Gauge pressure, it should remain less than 4 x 10-6 Torr while setting ​the Probe It should settle to around 2 x 10-6 or less when done. 
-The first check you need to make is on the vacuum status of the sourceThe source pressure should be less than 5x10<​sup>​-6</​sup>​ torr. If it is not find out why before going any further. +    - Raise the Probe Voltage to 3 KV as a startthis might be enough to get the plasma going
- +    ​- ​If there is still no plasma try raising the Mass Flow setting to as much as .5 SCCM to inspire ignition and then return to .35 SCCM.  
-If the control ​power for the machine, and/or for the source are off, or if the sodium canal supply is off or at zero current, you are starting from the cold state+    - Monitor ​the Probe Current, maintain ​mA +/- 1 mA, and continue raising the Probe Voltage until it reaches 6.5 KV.    
- +  - Set the Oven Voltage ​to 87 V, wait ten minutes. ​ After 10 minutes ​the Oven Temp should be around 150 C
-If the power to the source console is on and the sodium canal heater current is ~0.3 and a beam has been successfully extracted previouslythen the source is starting already warm+  - Set the Extractor Voltage to 5.5 KV and the Bias Voltage to a value that gives the desired injection energy ​less (6.5 + 5.5). Beam yield seems to increase with injection energy above 30 KeV.  Suggest trying 41 or 53 KeV settings if pulsing
- +    ​- ​
-If the power to the source console is on and the sodium canal heater current is ~0.55 A and there is ~0.75 A of arc and the extraction and canal supplies are onthen the source is in standby+
- +
-If the source ​is in any other state please ask before proceeding+
- +
-** Starting warm ** +
- +
-This applies when the source already been run successfully and the Sodium heater current left at ~0.A, and is the usual state from which to start the source+
-At the Control Console: +
- +
-1Check the source vacuum is less than 5x10<​sup>​-6</​sup>​ torr+
- +
-In the Vault: +
- +