Line 47: | Line 47: | ||
1. Turn source magnet up to ~0.2 A. (likely already set) | 1. Turn source magnet up to ~0.2 A. (likely already set) | ||
- | + | ||
2. Turn up the source gas until the pressure on the source tc gauge is ~150 microns. | 2. Turn up the source gas until the pressure on the source tc gauge is ~150 microns. | ||
- | + | ||
3. Set the Arc Voltage to 150 V. | 3. Set the Arc Voltage to 150 V. | ||
- | + | ||
4. Turn on the extraction supply, it should be set for 10-12 kV.� This will also bring up exchange canal power supply to about 13 kV. | 4. Turn on the extraction supply, it should be set for 10-12 kV.� This will also bring up exchange canal power supply to about 13 kV. | ||
- | + | ||
5. Turn on the focus supply, it should be set for ~2 kV. | 5. Turn on the focus supply, it should be set for ~2 kV. | ||
- | + | ||
6. Turn up the sodium heater current to 0.55 A, and the filament current up slowly, over 30 seconds, until the arc starts. | 6. Turn up the sodium heater current to 0.55 A, and the filament current up slowly, over 30 seconds, until the arc starts. | ||
- | + | ||
7. When the arc starts, 70 A for an old filament, 90 A for a new filament, observe the Arc Voltage and keep it at 150 V. | 7. When the arc starts, 70 A for an old filament, 90 A for a new filament, observe the Arc Voltage and keep it at 150 V. | ||
- | + | ||
8. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. | 8. Carry on increasing the filament current until the arc current reaches 2.5 A, keep the Arc Voltage at 150 V. | ||
- | + | ||
9. Increase the sodium heater current until it reaches 0.65 A, or whatever was used the last time the source was run – check the log. | 9. Increase the sodium heater current until it reaches 0.65 A, or whatever was used the last time the source was run – check the log. | ||
- | | + | |
10. Set the inflection magnet to 0.8 amps and LEFT deflection. | 10. Set the inflection magnet to 0.8 amps and LEFT deflection. | ||
- | | + | |
11. To obtain the maximum current adjust the following: | 11. To obtain the maximum current adjust the following: | ||
* Focus | * Focus | ||
Line 71: | Line 71: | ||
* Einzel lenses A and B | * Einzel lenses A and B | ||
* X and Y steerers | * X and Y steerers | ||
- | | + | |
12. Typical Settings: | 12. Typical Settings: | ||
* Extraction : 12 kV @ < 1 mA | * Extraction : 12 kV @ < 1 mA | ||
- | * Attachment: 13 kV @ 1 mA | + | * Attachment: 13 kV @ 1 mA |
* Focus: 2 kV @ 1.5 mA | * Focus: 2 kV @ 1.5 mA | ||
* Magnet : 0.2 A | * Magnet : 0.2 A | ||
Line 86: | Line 86: | ||
* 2Y: 5.00 | * 2Y: 5.00 | ||
* 2X: 5.00 | * 2X: 5.00 | ||
- | |||
Box Vacuum: 2.0x10E-6 torr | Box Vacuum: 2.0x10E-6 torr |