| Line 55: | Line 55: | ||
| b. Slowly raise the CONTROL GAIN to 2-3 for low current beams, around 5 for high current beams. | b. Slowly raise the CONTROL GAIN to 2-3 for low current beams, around 5 for high current beams. | ||
| - | c. Make fine adjustments to the TERMINAL VOLTAGE Dial to reach exactly the desired GVM value. | + | c. Make fine adjustments to the TERMINAL VOLTAGE Dial to reach exactly the desired GVM value. Very fine adjustments to the terminal voltage can be made using the FINE control located coaxial to the BIAS CURRENT control with a full range of 1.5 KV. |
| d. Switch OVERVOLTAGE PROTECTION to ENABLE. | d. Switch OVERVOLTAGE PROTECTION to ENABLE. | ||
| Line 79: | Line 79: | ||
| b. If not already visible on the control room computer screen, start the appropriate software GUI's using the " | b. If not already visible on the control room computer screen, start the appropriate software GUI's using the " | ||
| - | c. In the " | + | c. In the " |
| d. Run the slider to the desired current. | d. Run the slider to the desired current. | ||