| ====== Accelerator Start-up ====== | ====== Accelerator Start-up ====== | ||
| + | |||
| + | === ==== ** Clear all Areas ** ==== === | ||
| + | |||
| 1. **Verify Initial Machine Voltage setup:** | 1. **Verify Initial Machine Voltage setup:** | ||
| Line 27: | Line 30: | ||
| - | + | a. Switch the POWER SUPPLIES to ON, the yellow light should come on. Switch CHAIN MOTORS to ON, lights 1, 2, and 3 should all come on. Begin to increase the CHARGING VOLTAGE | |
| - | a. Switch the POWER SUPPLIES to ON, the yellow light should come on. Switch CHAIN MOTORS to ON, lights 1, 2, and 3 should all come on. Begin to increase the CHARGING VOLTAGE | + | **NOTE: If you will be raising the terminal voltage to 3.5MV or higher double click the TANDEMWATCH icon on the alphatross terminal monitor. |
| - | + | ** | |
| - | b. Continue to increase the CHARGING VOLTAGE until you reach the approximate TERMINAL VOLTAGE **__OR__** until the GRID VOLTAGE reaches | + | b. Continue to increase the CHARGING VOLTAGE until you reach the approximate TERMINAL VOLTAGE **__OR__** until you begin seeing any corona current. |
| Line 36: | Line 39: | ||
| **2. Corona Probe Adjustment** | **2. Corona Probe Adjustment** | ||
| - | FYI: The CORONA PROBE CURRENT will rise until it reaches the value set by the BIAS CURRENT. | + | After reaching the approximate desired terminal voltage with no corona current present, begin moving the corona points in (set corona current/ |
| a. Move the CORONA PROBE POSITION “IN” to make the GRID VOLTAGE more negative. | a. Move the CORONA PROBE POSITION “IN” to make the GRID VOLTAGE more negative. | ||
| Line 52: | Line 55: | ||
| b. Slowly raise the CONTROL GAIN to 2-3 for low current beams, around 5 for high current beams. | b. Slowly raise the CONTROL GAIN to 2-3 for low current beams, around 5 for high current beams. | ||
| - | c. Make fine adjustments to the TERMINAL VOLTAGE Dial to reach exactly the desired GVM value. | + | c. Make fine adjustments to the TERMINAL VOLTAGE Dial to reach exactly the desired GVM value. Very fine adjustments to the terminal voltage can be made using the FINE control located coaxial to the BIAS CURRENT control with a full range of 1.5 KV. |
| d. Switch OVERVOLTAGE PROTECTION to ENABLE. | d. Switch OVERVOLTAGE PROTECTION to ENABLE. | ||
| + | |||
| + | e. Do accelerator header and collect the NMR frequency and analyzer magnet current values. | ||
| **4. Beam Transport** | **4. Beam Transport** | ||
| - | a. Use the magnet predictor tool and set Quad 1 to the predicted value. | + | a. Use the magnet predictor tool and set Quad 1 to the predicted value. Do not make large abrupt changes to the Quad 1 current. |
| b. Tune beam and send it around the properly configured Analyzing Magnet. | b. Tune beam and send it around the properly configured Analyzing Magnet. | ||
| Line 67: | Line 72: | ||
| e. When in AUTO control the machine will go back to GVM control anytime the beam is interrupted. | e. When in AUTO control the machine will go back to GVM control anytime the beam is interrupted. | ||
| + | |||
| + | **5. Dipole Magnet Setup Instructions** | ||
| + | |||
| + | a. Confirm that the Heinzinger power supplies are powered ON in the Small Target Room with the following LED's illuminated: | ||
| + | |||
| + | b. If not already visible on the control room computer screen, start the appropriate software GUI's using the " | ||
| + | |||
| + | c. In the " | ||
| + | |||
| + | d. Run the slider to the desired current. | ||
| + | |||