playground:powerout
Differences
This shows you the differences between two versions of the page.
| Next revision | Previous revision | ||
| playground:powerout [2012/02/15 12:46] – created jacobs | playground:powerout [2026/06/25 16:12] (current) – external edit 127.0.0.1 | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| **Power Failure** | **Power Failure** | ||
| - | A power outage lasting longer than 1 minute | + | A power outage lasting longer than 3 minutes will begin to create adverse conditions that can not be recovered from without the attention of Accelerator staff and will take a minimum of 4 hours before operations can resume. Be Aware: |
| The Operator should take the following actions after power is restored: | The Operator should take the following actions after power is restored: | ||
| - | - Silence any alarms that have activated | + | 1) Silence any alarms that have activated, E1/E2, shutdown module, tritium monitor. |
| - | - Notify Accelerator Staff of the situation | + | |
| - | - Re-activate the cooling water loop pumps. The controls are located at the entry to the source area. (Ground the high voltage cage). Push each start button. The source cooling water must also be re-activated. This control is on the side of the cabinet that is against the wall by the sputter source. You may need to hold the “on” switch for several seconds before the pump will remain on. | + | |
| - | {{playground: | + | 2) Note how long the power was out. |
| - | - Check the status of the vault cryopump compressors. There is one in the high energy pumping station and one next to the low energy pumping station. If they are running, turn them off. Red toggle switch on the front of the HE compressor, yellow breaker on the rear of the LE compressor. | + | 3) Notify Accelerator Staff of the situation |
| - | {{playground: | + | 4) Check the cooling water loop pumps. The controls are located at the entry to the source area. (Ground the high voltage cage). If the green lights are still on no action is required. |
| - | - Proceed to the mechanical room. Stand at the yellow circle painted on the floor in front of the main breaker panel. Push the reset and start buttons on panels A, B, & C. This will activate 2 air handlers, and the water pump behind you. | + | {{playground: |
| - | {{playground: | + | 5) The source cooling water must also be re-activated. This control is on the side of the cabinet that is against the wall by the sputter source. You may need to hold the “on” switch for several seconds before the pump will remain on. |
| - | - Return to the Control Room | + | {{playground: |
| - | | + | |
| - | - On the control computer, go to the Restart tab and select | + | 6) Check the status of the vault cryopump compressors. There is one next to the LE and HE pumping stations. If they are running, turn them off. Yellow breaker on the rear of the LE compressor. |
| - | | + | |
| - | | + | |
| + | |||
| + | 7) No action is required in the upstairs mechanical room. HVAC systems should return to normal within 5 minutes of power restoration. | ||
| + | |||
| + | 8) If the accelerator was running when the power went out, set all of its controls | ||
| + | |||
| + | 9) Check the status of the magnets. Each magnet control module should be showing a red light for standby or off condition. | ||
| + | To restore to the console control_power (cups, steerers, chain drive etc.) you will have to go to the blue chain drive control box in the vault and press the “start” button. | ||
| + | |||
| + | 10) On the control computer, go to the Restart tab and select " | ||