| ===== NMR Setup ===== | ===== NMR Setup ===== | ||
| Line 5: | Line 4: | ||
| - Depress POWER switch if not illuminated. | - Depress POWER switch if not illuminated. | ||
| - After a 1 minute warm up period a horizontal line should appear on the oscilloscope face. | - After a 1 minute warm up period a horizontal line should appear on the oscilloscope face. | ||
| + | - Note: The oscilloscope now used is located on the console under the BPM oscilloscope (up and left of the NMR). | ||
| - You may have to adjust the VERTICAL POSITION control to find the horizontal line. | - You may have to adjust the VERTICAL POSITION control to find the horizontal line. | ||
| - Set the HORIZONTAL SWEEP to 60. | - Set the HORIZONTAL SWEEP to 60. | ||
| - | - Recycle the analyzing magnet if needed. | + | - Recycle the analyzing magnet if needed. |
| - Set the analyzing magnet power supply to a current slightly lower than the current predicted for this energy. | - Set the analyzing magnet power supply to a current slightly lower than the current predicted for this energy. | ||
| - | - Set the TUNING VERNIER control to about mid range on its 10 turn control. | + | - Set the TUNING VERNIER control to about mid range on its 10 turn control. |
| - Select a frequency about 1% less than the frequency desired for the beam energy. | - Select a frequency about 1% less than the frequency desired for the beam energy. | ||
| - In the light yellow (Tandem) window type the command " | - In the light yellow (Tandem) window type the command " | ||
| Line 16: | Line 16: | ||
| - The TANDEM window will inform you of how many steps needed to move approximately to the frequency you entered. | - The TANDEM window will inform you of how many steps needed to move approximately to the frequency you entered. | ||
| - When the NMR advances this number of steps it will inform you of the frequency the NMR has moved to at this step position. | - When the NMR advances this number of steps it will inform you of the frequency the NMR has moved to at this step position. | ||
| - | - The NMR frequency is read directly on the accelerator console by the Agilent 53131A Universal Counter located directly above the NMR. The counter that is read by the computer is on the left and is the standard counter to use for the NMR frequency. | + | - The NMR frequency is read directly on the accelerator console by the Agilent 53131A Universal Counter located directly above the NMR. The counter that is read by the computer is on the left and is the standard counter to use for the NMR frequency. |
| - If the counter is not updating the frequency readout you need to press the " | - If the counter is not updating the frequency readout you need to press the " | ||
| - If this frequency is not close enough to the desired frequency, repeat the nmrgoto command and it should get closer. | - If this frequency is not close enough to the desired frequency, repeat the nmrgoto command and it should get closer. | ||
| - Set the FIELD SWEEP control to the optimum position. | - Set the FIELD SWEEP control to the optimum position. | ||
| - set the VERTICAL GAIN control for approximately 1/16-inch of " | - set the VERTICAL GAIN control for approximately 1/16-inch of " | ||
| - | - Now that the NMR is set to a frequency slightly lower than the desired frequency, we can begin slowly increasing the analyzer magnet control upward toward the current that will show a resonance on the NMR. Typical resonance patterns and control positions can be seen in the {{HVEC_NMR.pdf|Original HVEC NMR Manual}} figure 1. The resonance will enter the display from the right side and move across the CRT display from right to left. When the resonance is found and centered on the CRT display, the magnet flux (frequency) can be read on the Agilent counter located above the NMR Fluxmeter. | + | - Now that the NMR is set to a frequency slightly lower than the desired frequency, we can begin slowly increasing the analyzer magnet control upward toward the current that will show a resonance on the NMR. Typical resonance patterns and control positions can be seen in the {{HVEC_NMR.pdf|Original HVEC NMR Manual}} figure 1. The resonance will enter the display from the right side and move across the CRT display from right to left. When the resonance is found and centered on the CRT display, the magnet flux (frequency) can be read on the Agilent counter located above the NMR Fluxmeter. |
| - The NMR Fluxmeter is a very sensitive instrument and you must proceed slowly when raising the analyzer magnet current or you can easily miss the resonance on the CRT. When the Analyzing Magnet computer control window is in the " | - The NMR Fluxmeter is a very sensitive instrument and you must proceed slowly when raising the analyzer magnet current or you can easily miss the resonance on the CRT. When the Analyzing Magnet computer control window is in the " | ||
| - Once the resonance is found on the CRT, slowly advance the magnet and NMR frequency by alternately increase the magnet a few clicks then raise the NMR frequency by issuing the command "up # | - Once the resonance is found on the CRT, slowly advance the magnet and NMR frequency by alternately increase the magnet a few clicks then raise the NMR frequency by issuing the command "up # | ||
| Line 30: | Line 30: | ||
| -- findnmr ; predict the NMR frequency from the analyzing magnet current and move the NMR to that frequency. | -- findnmr ; predict the NMR frequency from the analyzing magnet current and move the NMR to that frequency. | ||
| -- browse #steps ; move the NMR down #steps/2 then slowly step up #steps, then down #steps, repeating until enter is pressed. | -- browse #steps ; move the NMR down #steps/2 then slowly step up #steps, then down #steps, repeating until enter is pressed. | ||
| - | + | - INCREMENTAL ENERGY STEP. A couple of routines have been written to help tune the switcher magnet and quadrupole magnets to a new energy after a good beam tune has been established at the current energy and you wish to move to a new energy (especially if the energy move is small). This routine scales the magnet currents linearly based on the ratio of the previous energy analyzer current and the new analyzer current. To use this feature (after a good tune at one energy and before attempting to change to a new energy) type the command | |
| + | -- " | ||
| + | -- " | ||
| + | Note: It is possible using the up and down commands to move the NMR oscillator frequency past its minimum or maximum value. | ||
| + | |||
| + | {{HVEC_NMR.pdf|Original HVEC NMR Manual. Main frequency adjustment control and tracking feature now obsolete. | ||
| + | |||
| + | ===== NMR principal of operation explained on YouTube ===== | ||
| + | |||
| + | * [[https:// | ||
| + | |||
| + | ---- | ||
| + | |||
| + | {{nmr_panel.jpg}} | ||
| + | // | ||
| - | |||
| - | * {{HVEC_NMR.pdf|Original HVEC NMR Manual. Main frequency adjustment control and tracking feature now obsolete. | ||
| - | | ||