| - | * a. The digital display reads out the total integrated charge in Coulombs * Multiplier (including the decimal point). | + | {{bci.pdf|Beam Current Integrator writeup |
| - | | + | ---- |
| + | |||
| + | * a. The digital display reads out the total integrated charge in Coulombs * Multiplier (including the decimal point). | ||
| + | |||
| + | | ||
| * c. This is the instantaneous beam current value. | * c. This is the instantaneous beam current value. | ||
| Line 11: | Line 15: | ||
| * d. This switch is used to expand the full scale indication of the beam current on the meter only. It has no effect on the total integrated charge value. | * d. This switch is used to expand the full scale indication of the beam current on the meter only. It has no effect on the total integrated charge value. | ||
| * e. This switch selects the polarity of the input current. | * e. This switch selects the polarity of the input current. | ||
| + | |||
| * f. The integrate light is on when the Beam Current Integrator is enabled. | * f. The integrate light is on when the Beam Current Integrator is enabled. | ||
| + | |||
| * g. This switch turns the integrator on (switch to " | * g. This switch turns the integrator on (switch to " | ||
| + | |||
| * h. This pre-scaler control is used to set the decimal point in the digital display window. | * h. This pre-scaler control is used to set the decimal point in the digital display window. | ||
| + | |||
| * i. The Preset Charge switches set the total charge amount that will stop the integration and turn off the integrate control light when the digits match the digits in the display AND the Preset Charge (w.) switch is in the Enable position. | * i. The Preset Charge switches set the total charge amount that will stop the integration and turn off the integrate control light when the digits match the digits in the display AND the Preset Charge (w.) switch is in the Enable position. | ||
| + | |||
| * j. This control sets the minimum level of current needed to allow integration if the LD switch (q.) is up. This value is 0 - 100% of meter full scale. | * j. This control sets the minimum level of current needed to allow integration if the LD switch (q.) is up. This value is 0 - 100% of meter full scale. | ||
| + | |||
| * k. This green lamp is illuminated if the indicated beam current is within the LO limit control (j.) and the HI limit control (n.). If LD (q.) is up it will allow integrate control to be enabled. | * k. This green lamp is illuminated if the indicated beam current is within the LO limit control (j.) and the HI limit control (n.). If LD (q.) is up it will allow integrate control to be enabled. | ||
| + | |||
| * l. This meter indicates 0-100%. | * l. This meter indicates 0-100%. | ||
| + | |||
| * m. This red lamp is illuminated if the indicated beam current is not within the LO limit control (j.) and the HI limit control (n.). If LD (q.) is up it will inhibit integrate integration and integrate control will be off. | * m. This red lamp is illuminated if the indicated beam current is not within the LO limit control (j.) and the HI limit control (n.). If LD (q.) is up it will inhibit integrate integration and integrate control will be off. | ||
| + | |||
| * n. This control sets the maximum level of current allowed for integration if the LD switch (q.) is up. This value is 0 - 100% of meter | * n. This control sets the maximum level of current allowed for integration if the LD switch (q.) is up. This value is 0 - 100% of meter | ||
| full scale. | full scale. | ||
| + | |||
| * o. This switch will allow resets when up. If this switch is down, the front panel reset switch (p.) and external reset input will be ignored. | * o. This switch will allow resets when up. If this switch is down, the front panel reset switch (p.) and external reset input will be ignored. | ||
| + | |||
| * p. This switch when up will reset the Beam Current Integrator to 0 if the Reset Enable switch (o.) is up. No integration will take place until this switch is down. The external reset BNC will reset the Beam Current Integrator when a +5 volt level is applied to the external reset input AND the Reset Enable (o.) switch is up. | * p. This switch when up will reset the Beam Current Integrator to 0 if the Reset Enable switch (o.) is up. No integration will take place until this switch is down. The external reset BNC will reset the Beam Current Integrator when a +5 volt level is applied to the external reset input AND the Reset Enable (o.) switch is up. | ||
| + | |||
| * q. This switch will allow the Level Discriminators (j. and n.)to control the integrator when the switch is in the UP position. | * q. This switch will allow the Level Discriminators (j. and n.)to control the integrator when the switch is in the UP position. | ||
| + | |||
| * r. The Percent 0-100 meter reads BCI dead time in the DT (up) position and Beam Fluctuation (BF) in the down position. | * r. The Percent 0-100 meter reads BCI dead time in the DT (up) position and Beam Fluctuation (BF) in the down position. | ||
| + | |||
| * s. This is a SPARE (unused) switch. | * s. This is a SPARE (unused) switch. | ||
| + | |||
| * t. Fast (F) / Slow (S) meter response. | * t. Fast (F) / Slow (S) meter response. | ||
| + | |||
| * u. This is a SPARE (unused) switch. | * u. This is a SPARE (unused) switch. | ||
| + | |||
| * w. Preset Charge (Enable Disable). | * w. Preset Charge (Enable Disable). | ||
| - | * {{bci.pdf|Beam Current Integrator writeup by designer, Bob Young}} | + | // |