| ====== Oxygen Monitoring ====== | ====== Oxygen Monitoring ====== | ||
| + | ===== Hazards ===== | ||
| + | The OUAL has several locations that present an asphyxiation hazard due to oxygen depletion. | ||
| + | ==== SF6 ==== | ||
| + | The hazards from SF6 are located in 2 locations: | ||
| + | - The vault, and especially the pit | ||
| + | - The dock | ||
| + | |||
| + | === The Pit === | ||
| + | |||
| + | The accelerator tank is filled with SF6 at 480 to 650 kPa (70 to 90 PSI). There is a blower, some filters, some dryers, and some piping connected to the tank to keep the SF6 circulated, clean, and dry. It is possible for any of this piping or the tank itself to leak SF6. Since SF6 is significantly heavier than air, it will flow to the lowest point, and gather there, until mixing with the air and dissipating. | ||
| + | The pit under the accelerator is the lowest point in the lab. The O2 monitor can be seen from the top of the stairs leading into the pit. | ||
| + | A plot of the oxygen readings from the pit can be found at [[https:// | ||
| + | |||
| + | === The Dock === | ||
| + | |||
| + | The storage tank and compressor for the SF6 are located in the loading dock. The amount of SF6 stored in the tank is generally much less than what is stored in the accelerator tank. When the tank is open to atmosphere, all of the SF6 is in the storage tank on the dock. | ||
| + | The O2 monitor is located approximately 1.5m (5 ft) to the right of the door leading from the lab into the dock. | ||
| + | A plot of the oxygen readings from the dock can be found at [[https:// | ||
| + | |||
| + | ==== Lab 115 - LN2 ==== | ||
| + | The lab/office space in room 115 is where dewars of Liquid Nitrogen (LN2) are stored. | ||
| + | The O2 monitor in this room is located on the wall to the right of the door from the lab hallway, approximately 1m (3 ft) from the door. | ||
| + | |||
| + | ==== Monitoring Equipment ==== | ||
| + | |||
| + | >**Below this point is the nitty-gritty details on the oxygen monitors, wiring, manuals, etc. Operators shouldn' | ||
| + | |||
| + | === Equipment Info === | ||
| + | |||
| + | Oxygen deficiency is monitored in the lab using 2 [[https:// | ||
| + | |||
| + | Normal oxygen content is **20.9%**. | ||
| + | |||
| + | If low oxygen is detected, an alarm will sound, and the area should be evacuated, and accelerator staff contacted. | ||
| + | |||
| + | Correspondence with AMI can be found at: {{:: | ||
| + | |||
| + | * Sensor data sheet: | ||
| + | * Web link: [[https:// | ||
| + | * Local copy (retrieved 2021-06-22): | ||
| + | * Sensor (w/remote probe) data sheet: | ||
| + | * Web link: [[https:// | ||
| + | * Local copy (retrieved 2021-06-22): | ||
| + | * Operator manual: | ||
| + | * Web link: [[https:// | ||
| + | * Local copy (retrieved 2021-06-22): | ||
| + | |||
| + | == Sensor Info == | ||
| + | The monitors are equipped with remote probes, to allow placement closer to the floor, since the risk is primarily from heavier than air gasses. | ||
| + | |||
| + | = Calibration Info = | ||
| + | The sensors should be calibrated quarterly. | ||
| + | |||
| + | === Communication Info === | ||
| + | The oxygen monitors come equipped with 3 modes of external communication: | ||
| + | |||
| + | == RS-485 == | ||
| + | Communication with the RS-485 port on these devices is accomplished by using an FTDI USB-to-RS-485 adapter. | ||
| + | |||
| + | == 4-20 mA analog output == | ||
| + | These devices utilize a standard 4-20 mA analog signal for integration with control systems. | ||
| + | |||
| + | == Relay Outputs == | ||
| + | These units are equipped with 2 alarm output relays, each with NO and NC contacts. | ||
| + | |||
| + | === Data logging === | ||
| + | Data logging for these devices is accomplished by landing the 2-10V signal wires into the HP/ | ||